International Standards and Conformity Assessment for all electrical, electronic and related technologies

TC 66

Safety of measuring, control and laboratory equipment

 

Detail

Committee
Working Groups
Project Leader

Current

Status

Frcst Pub

Date

Stability

Date

TC 66MT 10 H. PufahlDELPUB2015-032018

History

Stage
Document
Downloads
Decision Date
Target Date
AMW
66/477/RR PDF file 57 kB
2012-06-28 
CCDV
66/482/CDV PDF file 152 kB
PDF file 168 kB
2012-06-292012-06
AFDIS
66/514/RVC Word file 114 kB
PDF file 152 kB
2013-09-022013-02
DECFDIS
2014-09-112014-02
RFDIS
2014-09-222014-09
CFDIS
66/552/FDIS

2014-11-282014-12
APUB
66/567/RVD PDF file 120 kB
2015-02-132015-01
BPUB
2015-02-162015-02
PPUB
2015-03-102015-03
DELPUB
2018-10-052017-06
WPUB
  

Abstract

IEC 61010-2-051:2015 is applicable to electrically operated laboratory equipment and its accessories for mechanical mixing and stirring, where mechanical energy influences the shape or size or homogeneity of materials and their accessories. Such devices may contain heating elements. It has the status of a group safety publication in accordance with IEC Guide 104. This third edition cancels and replaces the second edition published in 2003. It constitutes a technical revision and includes the following change from the second edition:
- exclusion of equipment, whose size and weight make unintentional movement unlikely, from the drop test in Clause 8,
- added requirement for interlock systems containing electric/electronic or programmable components to Clause 15,
- notes have been re-phrased according to ISO/IEC Directives.

This publication is to be read in conjunction with IEC 61010-1:2010.

Project

IEC 61010-2-051:2015 ED3

Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-051: Particular requirements for laboratory equipment for mixing and stirring

 

 

Associated Documents:

66/627A/RR

PDF file 27 kB
66/563/RM

PDF file 284 kB