International Standards and Conformity Assessment for all electrical, electronic and related technologies

TC 86

Fibre optics

 

Detail

Committee
Working Groups
Project Leader

Current

Status

Frcst Pub

Date

Stability

Date

TC 86WG 4 Hisashi IzumitaPPUB2015-122020

History

Stage
Document
Downloads
Decision Date
Target Date
AMW
86/416/RR PDF file 57 kB
2011-12-09 
1CD
86/417/CD PDF file 850 kB
2011-12-162011-12
CDM
86/425/CC Word file 179 kB
PDF file 196 kB
2012-05-182012-05
A2CD
86/425A/CC Word file 194 kB
PDF file 253 kB
2013-01-252012-12
2CD
86/460/CD PDF file 976 kB
2013-07-192013-02
A3CD
86/463/CC Word file 195 kB
PDF file 242 kB
2013-11-292013-11
3CD
86/468/CD PDF file 519 kB
2014-06-202014-02
ACDV
86/476/CC Word file 267 kB
PDF file 92 kB
2014-11-212014-10
CCDV
86/477/CDV
PDF file 789 kB
2015-02-202015-01
APUB
86/483/RVC Word file 250 kB
PDF file 170 kB
2015-06-192015-08
DECPUB
2015-10-132015-09
BPUB
2015-10-262015-10
RFDIS
2015-11-26 
PPUB
2016-01-132016-02
  

Abstract

IEC 62129-1:2016 specifies procedures for calibrating an optical spectrum analyzer that is developed for use in fibre-optic communications and designed to measure the power distribution of an optical spectrum. It does not apply to an optical wavelength meter that measures only centre wavelengths, a Fabry-Perot interferometer or a monochromator that has no display unit. This first edition of IEC 62129-1 cancels and replaces the first edition of IEC 62129, published in 2006. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
- update of term and definitions;
- update of calibration conditions;
- calculation change of uncertainties related to wavelength temperature dependence, power linearity, power level temperature dependence;
- move of Annex E to the bibliography.
Keywords: calibrating an optical spectrum analyser

Project

IEC 62129-1:2016 ED1

Calibration of wavelength/optical frequency measurement instruments - Part 1: Optical spectrum analyzers

 

 

Associated Documents:

86/418A/RM

PDF file 275 kB